Targeting

_images/TargetingApp.png

MaxQuant.Live enables global targeting of more than 25,000 peptides Christoph Wichmann, Florian Meier, Sebastian Virreira Winter, Andreas-David Brunner, Juergen Cox, Mat-thias Mann doi: 1101/443838

The Targeting app allows the realization of various targeting strategies within on user interface. First, the list of targeting peptides has to be populated and afterward the targeting strategy has to be defined and the parameters have to be set.

Steps to setup a targeting run

  1. Populate the peptide list
  2. Set global parameters
  3. Set up targeting action

Peptide ions list

The targeting should comprise a set of peptides of interest and a set of high abundant peptides used for the real-time corrections of mass-to-charge values, expected retention times and peptide intensities.

Column Description  
Modified sequence String containing the modified sequence optional; Needed if isotopic labeling is used
Mass Ion mass mandatory
Charge Ion charge state mandatory
Retention time Expected retention time (min) mandatory
Apex Intensity Expected Apex intensity optional
Fragments mz mz values of expected fragment ions optional; needed for ms2 based corrections
MaxIt Maximum injection time for triggered scans optional; if empty the global MaxIt is used
NCE factors Scaling factors for stepped CE (white space separated: 1.2 2.3) optional;
Collision Energies Collision energies for fragmentation (white space separated: 11 21) optional; if empty the global NCE is used
RealtimeCorrection Use this ion to calibrate the realtime correction  
TargetedMs2 Trigger Ms2 scan(s) if ion detected (see TargetedMs2 )  
TargetLabeled Use labeled counterpart of this ion as trigger ion (see TargetLabeled)  
TgtMultiInjection Trigger multi injection scans for this ion and the labeled counterparts  
TopNExclusion Exclude this ion from the stochastic Top-N method  

Mandatory information

Mandatory columns have a white background and columns containing optional information/settings have a gray background. The required information of every peptide ion is the mass, the expected retention time and its charge state.

Additionally the modified sequence should be provided to be able to identify the peptide later in the MaxQuant.Live log file. If the selected targeting strategy comprises the usage of heavy labels, the modified is used by MaxQuant.Live to calculate the mass differences between the light and heavy channel.

The Apex Intensity is optional. If not set, the ion counts as detected if the number of set isotopic if present in a survey scan. If set, the ion counts as detected if the ratio of the intensity of the most intense isotopic peak present in the survey scan and expected Apex intensity is higher than the IntensityPeakRatioThreshold.

The other columns contain checkboxes to enable actions called when the peptide is detected. These actions are also contained in the left list in the targeting app window.

Automatic population of the targeting table

The peptide table can either be filled out manually, or populated automatically by either pasting a table from the clipboard, or by using tab separated text files.

When pasting from the clipboard, a tabular calculation program like Microsoft Excel can be used to set up the peptide list. Select the rows of peptides in Excel, copy to the clipboard and click on the paste icon in MaxQuant.Live to paste into the peptide list. Make sure that the header row is also copied and that it contains the same labels as the MaxQuant.Live peptide list. A dialog window helps you to map the column headers of the copied content to the MaxQuant.Live table.

The option to populate the table by using tab separated files is the easiest one. First, the evidence files - which is part of the MaxQuant analysis output - is selected. The column headers mapping is detected automatically, but can be changed by the user. Please note, that the evidence table should be inspected beforehand and peptides without intensities sorted out. Of course, the selected file doesn’t need to be an evidence file but can be any tab separated text file containing the required information and the corresponding column headers.

To add the values of the Apex intensity and the fragment ions, which are not present in the evidence table, the msms table - also part of the MaxQuant analysis output - can be read in to complete the targeting list. As before, the column mapping can be checked by the user, before the files are parsed by the software and the MaxQuant.Live targeting list is populated accordingly.

Global settings

The left list consists of global settings as well as the settings of the various scan actions which can be activated for every peptide ion in the list by settings the corresponding checkboxes. Whe a list entry is selected the settings are shown in the box below.

Setup of targeting actions

There are three ways to select targeting actions for peptides in the list 1. Click on the checkbox for every peptide manually 2. Select one or more peptides in the list, then select an action from the list of actions in the left panel and click on the check mark symbol in the middle. 3. Add a column with the name of the action to your peptide list in Excel and will it with the Boolean values TRUE or FALSE. When pasting the peptide in MaxQuant.Live, the software will set the checkboxes automatically.

Global settings: SurveyScan

The survey scans for the peptide recognition can be either MS1 scan or BoxCar scan. The parameters are described in the BoxCar section.

Global settings: TopN

The TopN strategy is applied on top of the targeting scans. Note, that peptides can also be excluded from being selected as precursors by using the “ExcludeFromTopN” feature.

RealtimeCorrection

The adaptive real-time correction adjusts the expected retention times based on the deviations of the peptide observed before. We refer to our MaxQuant.Live paper for more information about the real-time correction algorithm.

Parameter Description Default value
MzTolerances Lower and upper bounds for the peptide mz tolerance [ppm] (4.5,10)
RetentionTimeTolerances Lower and upper bounds for the retention time tolerance[min] (1,10)
SigmaScaleFactorRt Scaling factor for retention time window adjustment 3
PeptideHistoryLength Time span considered to calculate the corrections [min] 2
MinUsedCorrectionPeptides Minimum number of peptides used to calculate the corrections 50
IntPeakRatioThreshold The ratio of the current peak intensity to the apex value has to be higher than this threshold 1.00E-05
PepdideDetectionIsoPeaks Number of isotopic peaks that need to be found 2
IsotopeTolerance Isotopic peaks distance tolerance [ppm] 9
Ms2DetectionNeeded Use only peptide those peptides for the realtime correction whose fragment ion have been already detected. Make sure that the targetedMs2 is activated and the Fragments mz column populated for these peptide ions. False
Ms2ExcludeDetectedPeps Disable re-fragmentation of Ms2-detected peptides False
Ms2NormIntensity Minimal valid intensity normalized to the max. Ms2 peak 0.1
Ms2MzTolerance Mz tolerance for fragment ion detection in Ms2 scans [ppm] 20

TargetedMs2

Parameter Description Default value
BatMode Fragment ion in every cycle within the retention time, even if not detected FALSE
AutoPriority Prioritize MS2 scan, if the peptide was not fragmented before TRUE
MaxNumOfScans Upper limit for the number of fragmentations per peptide. 0=no limit 0
MzOffset Shift the isolation window by this mz value [th] 0
LowerMzBound Lower mz boundary for Ms2 scans 100
CollisionEnergy Collision energy for fragmentation 27
LifeTime Max time span before a scheduled MS2 scan get deleted [ms] 2000
MaxIT Maximum ion injection time (ms) 28
Resolution MS resolving power at m/z 200 15000
AgcTarget AGC target value (charges) 100000
PositiveMode Ion polarity TRUE

TargetLabeled

This action adds the SILAC label masses to the peptides masses in the list. The “Modified is used to calculate the number Arg and Lys amino acids. The masses of SILAC labels can be freely defined here.

TargetedMultiInjection

With this action, scans are triggered in which the heavy and the light channel are both selected by the quadrupole and analyzed together. If fragmentation is switched on, this resembles a PRM experiment. If fragmentation is turned off, it corresponds to an SIM experiment. The pmSIM acquisition strategy described in the paper can be implemented with this action turning on the BatMode switch.

Parameter Description Default value
WindowSize Size of the isolation windows [th] 1.4
WindowOffset Shift the isolation window centers by this offset [th] 0
BatMode Fragment ion in every cycle within the retention time tolerance window, even if not detected FALSE
NceLight / Heavy Normalized collision energy for the light/heavy channel 0
ItBoxSize Maximum injection time per light or heavy mass range. If smaller than MaxIt, the MaxIt is spread over multiple injections, each fulfilling the max. injection time per box condition. 100
Isotopic label Masses of SILAC labels  
PositiveMode Ion polarity TRUE
CollisionEnergy Collision energy for fragmentation 0
LowerMzBound Lower mz boundary for Ms2 scans 100
LifeTime Max time span before a scheduled MS2 scan get deleted [ms] 1000
Resolution MS resolving power at m/z 200 30000
AutoPriority Prioritize MS2 scan, if the peptide was not fragmented before TRUE
MaxNumOfScans Upper limit for the number of fragmentations per peptide. 0=no limit 0
MaxIT Maximum ion injection time (ms) [not used] 50
MaxItLight /Heavy Maximum ion injection time for light/heavy channel (ms) 100
AgcTarget AGC target value (charges) 100000
InjectTargetL/H AGC target value for light/heavy peptide (charges) 50000

ExcludeFromTopN

Parameter Description Default value
BatMode Exclude ion from TopN within the retention time tolerance window , even if not detected FALSE
StaticExclusionTime Time span the peptide is excluded from TopN (30) 30
AdaptiveExclusionTime Automatic exclusion of the peptide from TopN TRUE